Acquidata Uromac Petite

The Uromac Petite data processor, transducers and Powerbook® computer can be disassembled quickly and easily for travel to multiple sites, providing a very flexible and portable urodynamics system.

Recording parameters:

Petite has a four input channel AcquiAmplifier and eight trace Acquidata display which provides measurement and display of:
  • Volume infused (Vfill)
  • Vesical pressure (Pves)
  • Abdominal pressure (Pabd)
  • Detrusor pressure (Pdet)
  • Voided flow (Qura)
  • Volume voided (Vvoid)
    - as numeric value in the report
  • On-screen filling rate window

Key features include:

  • UroReport patient report with P-Q plot and nomograms

  • User reconfigurable display including split screen (for simultaneous comparison of flow patterns)

  • Export of report and all waveform data to spreadsheets or databases

  • All raw data saved with each file can be retrospectively reviewed using any desired settings or off-line calculations

  • Simple fast two step re-calibration of transducers

  • Supplied complete with all transducers and catheters ready to start testing (user to supply bladder filling media)

  • Remote controller with 3m cable to allow single key operation of most functions

  • AcquiPole mounted Apple Stylewriter 1200® greyscale printer or remotely located HP Deskwriter® colour printer

Acquidata Specifications

  • All scale displays on screen are user adjustable to any desired range during and after acquisition of data. A set default button will reconfigure each channel to a standard setting.

  • Vfill: continuously adjustable range from 0 to over 2000ml

  • Pves/Pabd/Pdet: continuously adjustable range from -100 to over 500cm H20

  • Qvoid: continuously adjustable range from 0 to over 100ml sec-1

  • Vvoid: volume actually voided is printed into the final report

  • 10 preset and 20 user editable event comment keys.

  • Screen scrolling speed from 1mm sec-1 to 1cm sec-1

  • Sample rates adjustable from 20 to 500 samples/division.

  • Input signal conversion 12-bit resolution.

  • All channel zero balance button on both AcquiAmplifier and keyboard.

  • Screen selectable single channel zero balance.

  • Unlimited settings configuration files can be set and saved.

  • Archive storage - up to 40 patient tests per standard 1.44Mb diskette.

  • Patient isolation to class CF (cardiac protected).

  • Acquidata Uromac systems are designed to comply with the IEC 601.1 standard for electromedical equipment.